80I-110S Datasheet

DanaherTest Accessories

AC/DC Current Clamp 100A, 100kHz Bandwidth, 600V

The Fluke 80i-110s is a clamp-on AC/DC current probe designed to reproduce current waveforms in modern commercial and industrial power distribution systems. It is compatible with millivolt-measuring instruments like oscilloscopes, ScopeMeters, and power harmonic analyzers. The probe is optimized for accurate AC, DC, and AC+DC current measurements up to 100 A peak across electrical, electronic, and automotive applications.

Download PDF

80I-110S Key Specifications

Bandwidth
100kHz
Current Rating
100A
Voltage Rating (AC)
600V
Max Operating Temperature
50°C
Min Operating Temperature
0°C
Approvals
UL, CSA, IEC
Battery Type
9V
Height
36mm

Characteristics (datasheet)

Current Ranges
0 to 10A dc or ac peak; 0 to 100A dc or ac peak
Output Signals
10A range: 100 mV/A; 100A range: 10 mV/A
Basic Accuracy (DC to 1kHz)
0 to 10A: <3% of reading +50 mA (100 mV/A); 0 to 40A: <4% of reading +50 mA (10 mV/A); 40 to 80A: <12% of reading +50 mA (10 mV/A); 80 to 100A: <15% of reading (10 mV/A)
Useful Bandwidth (-3 dB)
0 to 100 kHz
Rise or Fall Time
<4 Psec.
Input Load Impedance
>1 M: in parallel with up to 100 pF.
Output noise level
10 mV/A typ. 480 PV pk-pk; 100 mV/A typ. 3 mV pk-pk
Temperature coefficient
2000 ppm/qC max. for temperature from 0 to 50 qC (32 to 132 qF)
Battery
9 volt battery, type IEC 6LR61

Absolute Maximum Ratings

Working Voltage (Clamps jaws to Ground)
600V ac rms on Installation Category II per IEC 1010-1 circuits; 300V ac rms on Installation Category III per IEC 1010-1 circuits
Floating Voltage (Output cable and connector to Ground)
600V ac rms on Installation Category II per IEC 1010-1 circuits; 300V ac rms on Installation Category III per IEC 1010-1 circuits
Max. non-destructive current (0 to 2 kHz)
140A peak
Max. non-destructive current (2 to 10 kHz)
110A peak
Max. non-destructive current (10 to 20 kHz)
70 A peak
Max. non-destructive current (20 to 50 kHz)
30A peak
Max. non-destructive current (50 to 100 kHz)
20A peak
Maximum safe circuit rating (CAT II / CAT III)
600V in Installation Category II (CAT II) or 300V in Installation Category III (CAT III)

80I-110S Applications

  • Electrical, Electronic and Automotive applications
  • commercial and industrial power distribution systems
  • electronic motor drives and ignition systems

80I-110S Pinout & Package

Package/Case
100
Mount
Clamp

The full pin configuration and pin descriptions are in the 80I-110S PDF datasheet above.

Other Danaher parts

  • FLUKE-561 — IR Digital Thermometer, -40°C to 1022°C, 1% Acc
  • FLUKE-28IIEX/ETL — TRMS MULTIMETER INTRINSICALLY SAFE VERSION OF 28II
  • C116 — Instrument Protection and Storage SOFT CARRYING CASE
  • FLUKE-700P09EX — Calibration Equipment I.S. PRESSURE MODULE 1500 PSIG
  • FLUKE-576-CF — THRM PHTO IR W/LOGNG CLSE FOCUS
  • 686527 — Fuse 250 V/125 mA, 686527, Fluke
  • PWS4205 — Linear Benchtop Power Supply, 20V 5A, USB Interface, 0-40°C
  • FLUKE-700PCK — Calibration Equipment PRESSURE MODULE CALIBRATION KIT

More Test Accessories

  • 930126100 — HIRSCHMANN TEST AND MEASUREMENT 930126100 CROCODILE CLIP, 6MM, GREY/BLACK, 6...
  • 973368100 — HIRSCHMANN TEST AND MEASUREMENT 973368100 PROBE, TEST, BLACK, 30VAC/60VDC, 4...
  • 972337002 — HIRSCHMANN TEST AND MEASUREMENT 972337002 TEST LEAD SET, BLK, RED, 1M, 1KV
  • 934066103 — HIRSCHMANN TEST AND MEASUREMENT 934066103 TEST LEAD, YELLOW, 2M, 60V, 32A
  • TA044 — PICO TECHNOLOGY TA044 Oscilloscope Probe, 70 MHz, High-Volta...
  • 934061709 — HIRSCHMANN TEST AND MEASUREMENT 934061709 Test Lead, 4mm Banana Plug to 4mm...
  • 4648-C-48 — Test Leads BNC TO MAXIGRABBE
  • SPC12055 — Connector assemblies, Test; Lead Length: 1ft; Insulator Color: Black; Cable Leng...

80I-110S FAQ

What is 80I-110S?

80I-110S is a Test Accessories from Danaher. AC/DC Current Clamp 100A, 100kHz Bandwidth, 600V.

Who manufactures 80I-110S?

80I-110S is manufactured by Danaher.

What package and pinout does 80I-110S have?

80I-110S comes in 100 package, clamp mounting. The pin configuration and pin descriptions are given in the PDF datasheet.

What is the operating temperature range of 80I-110S?

80I-110S operates from 0°C to 50°C.

Where can I download the 80I-110S datasheet PDF?

The 80I-110S datasheet PDF (452 KB) can be downloaded free from this page using the Download PDF button.

Specifications

Manufacturer
Danaher
File Size
452 KB

Ask AI Assistant

Ask questions about 80I-110S specifications

AI Powered

Click to ask a quick question:

Parameters

ApprovalsUL, CSA, IEC
Bandwidth100kHz
Battery Type9V
Current Rating100A
Height36mm
Length231mm
Max Operating Temperature50°C
Min Operating Temperature0°C
MountClamp
Package/Case100
Package Quantity1
Reach SVHC CompliantNo
RoHSCompliant
Voltage Rating (AC)600V
Weight330g
Width67mm