2001-SCAN Datasheet

KeithleyTest Equipment

KEITHLEY 2001-SCAN SCANNER CARD, 10 CHANNEL

Download PDF

Characteristics (datasheet)

Description
10-channel Scanner Card with two high-speed channels
Warranty Extension Option
2001-SCAN-3Y-EW
DC Voltage Range
1nV to 1100V
AC Voltage Range
100nV to 775Vrms
DC Current Range
10pA to 2.1A
DC In-Circuit Current Range
100µA to 12A
AC Current Range
100pA to 2.1A
Frequency Range
1Hz to 15MHz
Frequency Sensitivity
60mV to 775Vrms
Period Range
67ns to 1s
Crest Factor Range
20Hz to 1MHz
Peak Spikes–Repetitive
500ns to 16s
Peak Spikes–Single
2µs to 16s
Resistance 2-Wire Range
100nΩ to 1GΩ
Resistance 4-Wire Range
100nΩ to 2.1MΩ
Temperature–RTD Range
–200° to 630°C
Temperature–TC Range
–200° to 1820°C
DCV Range (200 mV) Full Scale
±210.00000 mV
DCV Range (2 V) Full Scale
±2.1000000 V
DCV Range (20 V) Full Scale
±21.000000 V
DCV Range (200 V) Full Scale
±210.00000 V
DCV Range (1000 V) Full Scale
±1100.0000 V
DCV Default Resolution (200 mV to 1000 V)
10 nV, 100 nV, 1 µV, 10 µV, 100 µV
DCV Input Resolution (200 mV to 1000 V)
100 nV, 1 µV, 10 µV, 100 µV, 1 mV
DCV Input Resistance (200 mV, 2 V, 20 V)
>10 GΩ
DCV Input Resistance (200 V, 1000 V)
10 MΩ ±1%
DCV Accuracy (5 Minutes, 200 mV to 1000 V)
3+3, 2+1.5, 2+1.5, 2+1.5, 10+1.5 ppm
DCV Accuracy (24 Hours, 200 mV to 1000 V)
10+6, 7+2, 7+4, 13+3, 17+6 ppm
DCV Accuracy (90 Days, 200 mV to 1000 V)
25+6, 18+2, 18+4, 27+3, 31+6 ppm
DCV Accuracy (1 Day, 200 mV to 1000 V)
37+6, 25+2, 24+4, 38+3, 41+6 ppm
DCV Accuracy (2 Years, 200 mV to 1000 V)
50+6, 32+2, 32+4, 52+3, 55+6 ppm
ACV Accuracy (20–50Hz, 200 mV to 750 V)
0.25 + 0.015%
ACV Accuracy (50–100Hz, 200 mV to 750 V)
0.07 + 0.015%, 0.1 + 0.015% (for 750 V)
ACV Accuracy (0.1–2kHz, 200 mV to 750 V)
0.03 + 0.015% (for 200 mV, 2 V), 0.04 + 0.015% (for 20 V, 200 V), 0.08 + 0.015% (for 750 V)
ACV Accuracy (2–10kHz, 200 mV to 750 V)
0.03 + 0.015% (for 200 mV, 2 V), 0.06 + 0.015% (for 20 V, 200 V), 0.09 + 0.015% (for 750 V)
ACV Accuracy (10–30kHz, 200 mV to 750 V)
0.035 + 0.015% (for 200 mV, 2 V), 0.08 + 0.015% (for 20 V, 200 V), 0.12 + 0.015% (for 750 V)
ACV Accuracy (30–50kHz, 200 mV to 750 V)
0.05 + 0.015%, 0.15 + 0.015% (for 750 V, typical)
ACV Accuracy (50–100kHz, 200 mV to 750 V)
0.3 + 0.015%, 0.5 + 0.015% (for 750 V, typical)
ACV Accuracy (100–200kHz, 200 mV to 200 V)
0.75 + 0.025%
OHMS Range (20 Ω, 200 Ω, 2 kΩ) Full Scale
21.000000 Ω, 210.00000 Ω, 2100.0000 kΩ
OHMS Default Resolution (20 Ω, 200 Ω, 2 kΩ)
1 µΩ, 10 µΩ, 100 µΩ
OHMS Resolution (20 Ω, 200 Ω, 2 kΩ)
10 µΩ, 100 µΩ, 1 mΩ
OHMS Current Source (20 Ω, 200 Ω, 2 kΩ)
9.2 mA, 0.98 mA, 0.98 mA
OHMS Accuracy (24 Hours, 20 Ω to 2 kΩ)
29 + 7, 24 + 7, 22 + 4 ppm
OHMS Accuracy (90 Days, 20 Ω to 2 kΩ)
52 + 7, 36 + 7, 33 + 4 ppm
OHMS Accuracy (1 Year, 20 Ω to 2 kΩ)
72 + 7, 56 + 7, 50 + 4 ppm
OHMS Accuracy (2 Years, 20 Ω to 2 kΩ)
110 + 7, 90 + 7, 80 + 4.5 ppm

2001-SCAN Pinout & Package

Mount
Cable

The full pin configuration and pin descriptions are in the 2001-SCAN PDF datasheet above.

Other Keithley parts

  • 2750/E — KEITHLEY 2750/E Bench Digital Multimeter, Current, Voltage,...
  • 7001 — Digital Multimeters Switch System
  • 2001-TCSCAN — KEITHLEY 2001-TCSCAN SCANNER CARD, THERMOCOUPLE, 10 CH
  • 6521 — SCANNER CARD; Accessory Type:Scanner Card; For Use With:Keithley 6517B Electrome...
  • 6220/2182A — Digital Multimeters Delta Mode System
  • 7158 — Digital Multimeters Low Current Scanner Card
  • 2182A — Digital Multimeters Nanovoltmeter
  • 2636A — METER, SOURCEMETER, DUAL, FA

More Test Equipment

  • 72-7595 — 4-IN-1 ANEMOMETER, HYGROMETER, LUXMETER, THERMOMETER; Air Velocity Range:-; Beau...
  • TEK-DPG — Tektronix Oscilloscope Module Deskew Pulse Generator TEK DPG, For Use With DPO/M...
  • 72-8705 — TENMA 72-8705 Oscilloscope, 2 Analogue, 50 MHz, 1 GSPS, 600 kpts, 7 ns
  • 18712 — Hearing u0026 Vision Aids System 250 FL Micro 120V
  • U1733P — LCR Meter Kit, U1733P, Keysight
  • 6613C-240-PLG — POWER SUPPLY, 50V, 1A 50W GPIB RS232
  • U1271A — Multimeter digital TRMS AC 30000 digits 1000 VAC 1000 VDC 10 ADC, U1271A, Keysig...
  • ACD-15PRO — Clamp Multimeters u0026 Accessories 2000A DIG CLAMP-ON MULTIMETER

2001-SCAN FAQ

What is 2001-SCAN?

2001-SCAN is a Test Equipment from Keithley. KEITHLEY 2001-SCAN SCANNER CARD, 10 CHANNEL.

Who manufactures 2001-SCAN?

2001-SCAN is manufactured by Keithley.

Where can I download the 2001-SCAN datasheet PDF?

The 2001-SCAN datasheet PDF (333 KB) can be downloaded free from this page using the Download PDF button.

Specifications

Manufacturer
Keithley
File Size
333 KB

Ask AI Assistant

Ask questions about 2001-SCAN specifications

AI Powered

Click to ask a quick question:

Parameters

MountCable
Package Quantity1
Reach SVHC CompliantUnknown
RoHSCompliant